AutoRead

Monitors the electrode drift and releases the measured
value only after the stability criterion has been reached. In
this way, this procedure ensures the highest degree of pre-
cision and reproducibility.

Calibration

The cell constant is determined through calibration. To do
so, the conductivity measuring cell is immersed into a series
of aqueous salt solutions with exactly known electric con-
ductiviy. The relevant conductivity values are determined
using the conductivity measuring instrument.

Cell constant

Linear factor describing the geometrical dimensions of a
measuring cell and its electrodes.

l

With the value of the cell constant you can roughly de-
scribe the application range of a conductivity measuring
cell.

l

The cell constant is determined through calibration in one
or several calibration solutions.

Conductivity

The conductivity value is a sum parameter for the ion con-
centration of a test sample.

Control standard

solution

Solution with a known conductivity to determine or check the
conductivity.

Drift control

See A

UTO

R

EAD

.

Reference

temperature

In order to compare values measured at different tempera-
tures, the values have to be converted to a fixed tempera-
ture. This temperature is 25 °C, or, as an exception, 20 °C.