What to do if...

143

6

What to do if...

Cause

Remedy

– Measuring cell not

connected

– Connect measuring

cell

– Cable broken

– Replace electrode

Cause

Remedy

– Measuring cell

contaminated

– Clean measuring cell;

if necessary, replace it

– Unsuitable calibration

solution

– Check calibration

solutions

Cause

Remedy

– Batteries almost

depleted

– Replace batteries

(see section 5.1
M

AINTENANCE

)

Cause

Remedy

– Operating state

undefined or EMC
electric stress unallowed

– Processor reset:

Press the <auto read>
key and switch on
instrument

Cause

Remedy

– Timeout of the interface

– Check connected

instrument

Error message,

I /

Error message,



LoBat

Instrument does not

react to keystroke

*(

display