Xylem CG 853P User Manual
Page 24

Operation
100
AutoRead AR
(Drift control)
The
AutoRead
function (drift control) checks the stability of
the measurement signal. The stability has a considerable
effect on the reproducibility of the measured value.
.
Note
The current AutoRead measurement (with acceptance of
the current value) can be terminated at any time by pressing
<run/enter>.
1 Call up the measuring mode by pressing <
ì
>.
2 Activate the AutoRead function by pressing
<auto read>.
The current measured value is frozen (Hold function).
3 Start the AutoRead function by pressing
<run/enter>.
AR flashes on the display until a stable measured val-
ue is reached.
This measured value is transmitted to the interface.
4 If necessary, start the next AutoRead measurement
by pressing <run/enter>.
5 To terminate the AutoRead function: Press the
<auto read> key.