Operation

19

ba75369e04

01/2011

4.2.2

Measuring the D. O. concentration

Note
When measuring the concentration of test samples with a salt content of
more than 1 g/l, a salinity correction is required. For this, you have to measure
and input the salinity of the measured medium first. This is described in sec-
tion 4.3.3 E

NTERING

THE

SALT

CONTENT

(

SALINITY

).

Before measuring the oxygen, you have to switch on the salt content correc-
tion (see below).

You can measure the oxygen content without a salt content correction as fol-
lows:

Switching on/off

the salt content

correction

Proceed as follows to switch on the salt content correction:

1

Perform the preparatory activities according to section 4.2.1.

2

Immerse the D. O. probe in the test sample.

3

Press the <M> key repeatedly until the D. O. concentration in mg/l
appears on the display.

ARng

%

LoBat

Cal

AR

8

°

F

Sal

mbar

S

1

RCL

Auto

Store

1/cm

/

K

%

TDS

UpH

Sal

mV/pH

S/

m

cm

cm

M

Lin

Tref25 Tref20

nLF

AutoCal DIN

AutoCal TEC

Time

Baud

Day.Month No.

Ident

Year

6

7

1

9 2

7

O

°

C

mg/l

TP

1

In the D. O. concentration measuring mode, switch on the salt con-
tent correction with <

>. The SAL display indicator appears on the

display. The value that was entered is taken into consideration during
the measurement (see section 4.3.3).

cm

cm

M

%

LoBat

Cal

AR

8

°

F

mbar

S

1

RCL

Auto

Store

1/cm

/

K

%

TDS

UpH

Sal

mV/pH

S/

m

Lin

Tref25 Tref20

nLF

AutoCal DIN

AutoCal TEC

Time

Baud

Day.Month No.

Ident

Year

6

7

1

4 2

5

O

°

C

mg/l

ARng

TP

Sal

Advertising