What to do if...

handylab Ox 12

50

ba75369e04

01/2011

Error message

E7

AR flashes

continuously

Measured value

too low

Measured value

too high

Display

to

Probe symbol

flashes

Cause

Remedy

Membrane damaged

– Membrane damaged

– Regenerate probe

– Membrane head not screwed

on tight enough

– Screw membrane head tight

Cause

Remedy

No stable measured value

– Membrane contaminated

– Clean membrane

Cause

Remedy

– Insufficient

flow

– Provide flow to the probe

Cause

Remedy

– High amount of dissolved

substances

– Correct solubility function using

the salinity equivalent

– Air bubbles bump on the

membrane with high velocity

– Avoid direct flow to the

membrane

– The carbon dioxide pressure

is too high (> 1 bar)

– Measuring not possible

Cause

Remedy

– Time-out of the interface

– Check the instrument that is

connected

Cause

Remedy

– Calibration interval expired

– Recalibrate the measuring

system