3 calibrating, Calibrating – Xylem handylab OX 12 User Manual

Page 23

Advertising
background image

handylab Ox 12

Operation

23

ba75369e04

01/2011

4.3

Calibrating

Why calibrate?

D. O. probes age. This changes the slope of the D. O. probe. Calibration de-
termines the current slope of the probe and stores this value in the instru-
ment.
Thus, you should calibrate at regular intervals.

Calibration

procedures

The handylab OX 12 provides 2 calibration procedures:

z Calibration in water vapor-saturated air.

Use an air calibration vessel for calibration.

z Calibration via a comparison measurement (e. g. Winkler titration accord-

ing to DIN EN 25813 or ISO 5813). At the same time, the relative slope is
adapted to the comparison measurement by a correction factor.

When to calibrate?

z After connecting another D. O. probe
z If the sensor symbol flashes, i. e. after the calibration interval has expired

Setting the

calibration interval

The calibration interval (Int 3) is set to 14 days in the factory. You can select
the interval in the range of 1 ... 999 days.

Printing the

calibration

record

The calibration protocol contains the calibration data of the current calibration.
You can transmit the calibration protocol to a printer via the serial interface
(see page 37).

Note
You can automatically print a calibration protocol after the calibration. To do
so, connect a printer to the interface according to section 4.5.3 before cali-
brating. After a valid calibration, the record is printed.

8

LoBat

Time

Baud

Day.Month No.

Ident

Year

Sal

%

mbar

1

S

RCL

AR

Store

Auto

Cal

°

F

1/cm

/

K

%

TDS

UpH

Sal

mV/pH

S/

m

cm

cm

M

Lin

Tref25 Tref20

nLF

AutoCal DIN

AutoCal TEC

6

7

1

9 2

7

O

°

C

mg/l

ARng

TP

Advertising