Selecting the measuring range, AutoRange

AutoRange

Each of the measuring parameters has 2 measuring ranges available. Au-
toRange
causes the measuring instrument to change to measuring range II
if measuring range I is exceeded, and also vice versa.

To switch the automatic measurement range selection on/off, see section
4.6 C

ONFIGURATION

. When the AutoRange function is switched on, ARng ap-

pears on the display.

4.2.6

AutoRead AR (drift control)

The AutoRead function (drift control) checks the stability of the measurement
signal. The stability has a considerable effect on the reproducibility of the
measured value.

1

Select the measuring mode with <M>.

2

Activate the AutoRead function with <AR>. The current measured
value is frozen (hold function).

3

Start AutoRead with <RUN/ENTER>.
The AR display indicator flashes until a stable measured value is
reached.

4

If necessary, start the next AutoRead measurement with <RUN/EN-
TER>
.

%

LoBat

Cal

AR

8

°

F

Sal

mbar

S

1

RCL

Auto

Store

1/cm

/

K

%

TDS

UpH

Sal

mV/pH

S/

m

cm

cm

M

Lin

Tref25 Tref20

nLF

AutoCal DIN

AutoCal TEC

Time

Baud

Day.Month No.

Ident

Year

6

7

1

9 2

7

O

°

C

mg/l

ARng

TP

7

RCL

%

LoBat

Cal

8

°

F

Sal

mbar

S

1

Auto

Store

1/cm

/

K

%

TDS

UpH

Sal

mV/pH

S/

m

cm

cm

M

Lin

Tref25 Tref20

nLF

AutoCal DIN

AutoCal TEC

Time

Baud

Day.Month No.

Ident

Year

6

7

1

7 0

O

°

C

mg/l

TP

ARng

AR

Advertising