Operation

25

ba75369e04

01/2011

4.3.1

Calibrating with the air calibration vessel

Proceed as follows to calibrate the instrument:

Note
The sponge in the air calibration vessel must be moist (not wet).

1

Connect the D. O. probe to the measuring instrument.

2

Keep an air calibration vessel ready for use.

3

Put the D. O. probe into the air calibration vessel.

4

Press the <CAL> key repeatedly until O

2

CAL appears.

5

Press the <RUN/ENTER> key. AutoRead is active, AR flashes.

6

As soon as a stable value is reached, the AR display indicator stops
flashing. This indicates that the calibration is complete; the sensor
symbol shows the relative slope that was determined and the sensor
evaluation (see page 24).