What to do if...
49
ba75369e04
01/2011
6
What to do if...
LoBat display
Instrument does
not react to key-
stroke
Error message
0FL
Error message
E3
Cause
Remedy
– Batteries almost empty
– Replace batteries
)
Cause
Remedy
– Operating condition
undefined or EMC load
unallowed
– Processor reset:
Press the <CAL> and
<ON/OFF> keys at the same
time and release them again.
The software version is
displayed.
Cause
Remedy
Display range exceeded
D. O. probe:
– Not connected
– Connect the probe
– Cable broken
– Replace probe
– Depleted
– Replace probe
– Short-circuit between gold
and lead electrode
– Clean probe and replace it if
necessary
Cause
Remedy
Invalid calibration
D. O. probe:
– Electrolyte solution depleted
– Regenerate probe
– Membrane contaminated
– Clean membrane
– Electrode system poisoned
– Regenerate probe
– Worn out
– Replace probe
– broken
– Replace probe