2 calibrating via a comparison measurement, Calibrating via a comparison measurement – Xylem handylab OX 12 User Manual

Page 27

Advertising
background image

handylab Ox 12

Operation

27

ba75369e04

01/2011

4.3.2

Calibrating via a comparison measurement

Proceed as follows to calibrate the instrument:

1

Connect the D. O. probe to the measuring instrument.

2

Press the <CAL> key repeatedly until O

2

FAC appears.

3

Immerse the D. O. probe in the calibrating solution.

4

Press the <RUN/ENTER> key.
The AutoRead measurement begins. If the measured value is stable,
the instrument displays the determined measured value and the cor-
rection factor of the relative slope (initial value = 1.000).

5

Remove the D. O. probe from the calibrating solution.

6

Perform a comparison measurement (determine the nominal value).
Leave the measuring instrument switched on until the nominal value
is determined.

7

Set the displayed value on the measuring instrument to the nominal
value with <

> <>. This changes the related correction factor of

the relative slope.