4.3.3

Entering the salt content (salinity)

A salt content correction is required in the oxygen concentration measure-
ment of test samples with a salt content of more than 1 g/l. For this, you have
to enter the salinity equivalent (the measured salinity) of the test sample
(range 0.0 - 70.0) and to switch on the salinity correction.

Entering the

salt content

Note
How to switch on the salt content correction is described on page 19.

Parameter

Value range

Salinity

0.0 ... 70.0 in steps of 0.1

1

Determining the salinity of the test sample (any method).

2

Press the <CAL> key repeatedly until Sal appears on the display.

3

Enter the salt content with <

> <>.

4

Switch to the measuring mode with <M>.