10 timing/misc setup – Xylem SDI-12 Data Logger iLink User Manual

Page 43

Advertising
background image

i

Quest (NZ) Ltd - PO Box 15169 HAMILTON Tel: +64 7 957 8160 - Fax: +64 7 957 8162 - Email:

iLink - Configuration and Support Utility

43

[email protected]

Maximum Raw

Applies to: Analogue Input

This setting indicates the maximum raw value that can be stored for this particular analogue input. It is
unlikely that this setting will need to be altered.

Minimum E.U

Applies to: Analogue Input

Values output by the sensor are converted to actual Engineering Units (E.U) by the logger program. This
setting specifies the minimum value in actual engineering units.

Maximum E.U

Applies to: Analogue Input

This setting specifies the maximum value in actual engineering units.

Logging Mult

Applies to: All Input Types
{1, 10, 100, 1000, 0.1, 0.01, 0.001}

The logging multiplier setting is used to convert the actual values (in Engineering Units) into integer values
for storage in the logger database. Thus in the screenshot above, in order to convert the E.U values of 0
– 5 into the stored values of 0 – 4095 a multiplier of 1000 is required. Other ranges of E.U values may
require different logging multipliers to bring them within the range 0 – 4095. Select the most appropriate
logging multiplier from the list.

E.g.

E.U Range: 0 – 1

Logging multiplier: 1000

E.U Range: 0 – 2

Logging multiplier: 1000

E.U Range: 0 – 10

Logging multiplier: 100

E.U Range: 1000 – 10000

Logging multiplier: 0.01


When working with unloaded logged data, the multiplier is used in reverse to return the logged values
back to their correct Engineering Units.

Scale Factor

Applies to: Counter, SDI-12, Sync serial, Up/Dn Encoder, Derived

This is used to specify the scale factor for digital devices, e.g. SDI-12.

6.3.1.10 Timing/Misc Setup


Warm Up

This setting is only applicable if the sensor is to be powered by the logger (using a Control Output). It
specifies the number of seconds to turn on the sensor before each reading is taken. If this value is zero,
the sensor will be permanently powered up.

Sampling

Enter the sampling frequency here. This setting specifies how often the logger will take a reading from the
sensor.

Logging

Enter the logging interval here. This setting specifies how often the DS-4483 will log a value for this
sensor. See also Event Mode (Timed) and Event Mode (Instant) below.



Advertising