2 inserting the battery, Inserting the battery – Xylem WQL-pH pH DATALOGGERS User Manual

Page 12

Advertising
background image

Commissioning

WQL-pH

86

ba75821e01

10/2009

3.1.2

Inserting the battery

Note
The logger is powered by a 3.6 V lithium battery, size AA, as provided in the
scope of the delivery (see chapter 13 T

ECHNICAL

DATA

WQL-Log). Other

batteries of the same size such as alkaline manganese batteries do not
provide the required operational voltage.
Operating time, see section 11.4 B

ATTERY

.

Fig. 3-1

Inserting the battery

1

Unscrew the cap (pos. 1 in Fig. 3-1).

2

Unscrew the sleeve (pos. 2).

3

Insert the battery When doing so, make sure that the battery is
correctly positioned (see label in the battery compartment).

4

Screw the sleeve on the shaft with the shorter thread. At the rear end
of the sleeve there is a thin O-ring inside (pos. 3). Make sure that this
O-ring is evenly positioned in the groove and is not twisted.

1

2

3

Advertising