Calibrating for pH measurements

WQL-pH

110

ba75821e01

10/2009

Calibration record

7

If you want to carry out a single-point calibration: Press Complete.
This completes the single-point calibration; the calibration record is
displayed (with the Nernst slope of -59.2 mV/pH at 25 °C).

Or continue with the two-point calibration:

8

Thoroughly rinse the logger and electrode with deionized water.

9

Immerse the logger in the second buffer solution.

10

Start the measurement with Continue.
The AutoRead running... note appears and the electrode voltage U
(mV) and temperature are displayed.
The calibration record with the value of the slope (mV/pH) and the
value of the asymmetry (mV) is displayed as soon as a stable value
is recognized.

11

Accept the calibration with Accept. The message Calibration
successful
appears and the new calibration will now be used for
measurement.
If you press Cancel, the logger discards the new calibration and will
continue to use the previous calibration values.

Note
The calibration line is determined by linear regression.

Note
You can view the calibration records at any time (see section 5.4 V

IEWING

THE

CALIBRATION

RECORDS

).