WQL-pH
Calibrating for pH measurements
111
ba75821e01
10/2009
Calibration
evaluation
After calibrating, the current system condition is evaluated. The asymmetry
and slope are evaluated separately. The probe symbol shows the worse case
evaluation of both.
.
Display
Asymmetry [mV]
Slope
[mV/pH]
-15 ... +15
-60.5 ... -58
-20 ... +20
-58 ... -57
-25 ... +25
-61 ... -60.5
or
-57 ... -56
Clean the electrode
according to the electrode
operating manual
-30 ... +30
-62 ... -61
or
-56 ... -50