3 calibrating with autocal, Calibrating with autocal – Xylem WQL-pH pH DATALOGGERS User Manual
Page 34
Calibrating for pH measurements
WQL-pH
108
ba75821e01
10/2009
5.3
Calibrating with AutoCal
Preparatory
activities
The following example covers the AutoCal calibration using WTW technical
buffer solutions.
Calibration
Note
Always check the selected buffer set before calibrating.
From the respective buffer set you can use any 2 buffers in any order.
Note
We recommend to always calibrate with the protective hood screwed on so
the electrode is protected. Use a stand as necessary.
1
Connect the WQL-pH logger (with installed electrode) to a USB
interface of your PC via the USB cable.
2
Keep the buffer solutions ready.
3
In the Device menu, select the menu item, Calibrate.
The calibration window appears.
4
Immerse the logger (with the electrode installed) in the first
buffer solution.
5
Use the Continue button to start the measurement.