Setting up a logging job - setting parameters

You can set the following parameters for a logging job (i.e. the job to
determine and record measured values according to your requirements over
a certain period of time):

Tab

Setting

Setting options
(default setting in bold)

Recording

Interval

z

1 second, 5 seconds,
10 seconds, 30 seconds

z

1 minute, 5 minutes, 10 minutes,
15 minutes, 30 minutes

z

1 hour, 2 hours, 3 hours,
6 hours, 9 hours, 12 hours,
24 hours

Start

z

Immediately
(when the setting mode is quit and the
logger is separated from the PC
program)

z

Logger key button
(when the key button at the control
panel of the logger is pressed)

Stop

z

Time period
(1 hour ... 365 days)

z

Memory full
(the loggers stops logging only when
the memory is full, i.e. when 600,000
datasets have been stored)

Measured
value ID

You can assign a measured value ID
(consisting of up to 20 alphanumerical
characters, no special characters) e.g.
the name of the measuring site).
Default setting: default

pH/mV

Measuring
mode

z

pH

z

mV

Time

Automatic
time
adjustment

z

Yes (date/time for the logger is
automatically synchronized with that
of the PC)

z

No
(date/time for the logger is set
manually)