Xylem WQL-pH pH DATALOGGERS User Manual

Page 15

Advertising
background image

WQL-pH

Commissioning

89

ba75821e01

10/2009

Note
The protection hood can be screwed on for transport or storage even if the
watering cap is plugged on. Please follow the notes on the storage of the
electrode in the operating manual of the electrode.

3

Pull out the blind plug (pos. 3).

4

Prepare the electrode (pos. 4):

– Make sure that both O-rings (pos. 5 and 6 ) are mounted, clean

and dry.

– Remove the watering cap (pos. 7) and dab dry the shaft of the

electrode with a clean paper towel.

– Position the coupling ring (pos. 2) on the shaft of the electrode.
– Plug on the watering cap (pos. 7) to protect the electrode for the

time of the installation.

5

Insert the thus prepared electrode (pos. 8) in the electrode
receptacle:

– Position the electrode at the opening and, against the pressure of

the air cushion inside, slowly push it in approx. 2 cm, until the air
escapes and the counter pressure is noticeably reduced.

– Then push in the electrode as far as it will go and turn it until the

electrical plug connection locks in place using light pressure. The
electrode is completely inserted when it cannot be turned any
more.

6

Screw on the coupling ring (pos. 2) by hand to the limit.

7

For calibrating and measuring remove the watering cap.

8

Screw on the protective hood (pos. 1 in Fig. 3-2).