5 ph calibration, Ph calibration – Xylem PH3210 FIELD PH METER User Manual

Page 34

Advertising
background image

Operation

pH 3210

34

ba75793e02

12/2008

4.4.5

pH calibration

Why calibrate?

pH combination electrodes age. This changes the zero point
(asymmetry) and slope of the pH combination electrode. As a result, an
inexact measured value is displayed. Calibration determines and
stores the current values of the zero point and slope of the electrode.
Thus, you should calibrate at regular intervals.

When do you have to

calibrate?

z After connecting another combination electrode

z When the calibration interval has expired

Buffer sets for

calibration

You can use the buffer sets quoted in the table for an automatic
calibration. The pH values are valid for the specified temperature
values. The temperature dependence of the pH values is taken into
consideration during the calibration.

No.

Buffer set*

pH

values

at

1

ConCal

Any

Any

2

NIST/DIN
DIN buffers according to DIN 19266
and NIST Traceable Buffers

1.679
4.006
6.865
9.180

12.454

25 °C

3

TEC
WTW Technical buffers

2.000
4.010
7.000

10.011

25 °C

4

Merck 1*

4.000
7.000
9.000

20°C

5

Merck 2 *

1.000
6.000
8.000

13.000

20°C

6

Merck 3 *

4.660
6.880
9.220

20°C