What to do if...

pH 3210

58

ba75793e02

12/2008

No stable measured

value

Sensor symbol flashes

Display

Cause

Remedy

pH combination electrode:

– Junction contaminated

– Clean junction

– Membrane contaminated

– Clean membrane

Test sample:

– pH value not stable

– Measure with air excluded if

necessary

– Temperature not stable

– Adjust temperature if

necessary

Combination electrode + test
sample:

– Conductivity too low

– Use suitable combination

electrode

– Temperature too high

– Use suitable combination

electrode

– Organic liquids

– Use suitable combination

electrode

Cause

Remedy

– Calibration interval expired

– Recalibrate the measuring

system

Cause

Remedy

– batteries almost empty

– Replace the batteries (see

section 5.1 M

AINTENANCE

)