What to do if...
pH 3210
58
ba75793e02
12/2008
No stable measured
value
Sensor symbol flashes
Display
Cause
Remedy
pH combination electrode:
– Junction contaminated
– Clean junction
– Membrane contaminated
– Clean membrane
Test sample:
– pH value not stable
– Measure with air excluded if
necessary
– Temperature not stable
– Adjust temperature if
necessary
Combination electrode + test
sample:
– Conductivity too low
– Use suitable combination
electrode
– Temperature too high
– Use suitable combination
electrode
– Organic liquids
– Use suitable combination
electrode
Cause
Remedy
– Calibration interval expired
– Recalibrate the measuring
system
Cause
Remedy
– batteries almost empty
– Replace the batteries (see
)