pH 3210
Operation
45
ba75793e02
12/2008
Note
For single-point calibration, the instrument uses the Nernst slope
(-59.16 mV/pH at 25 °C) and determines the zero point of the
electrode.
7
Wait for the measurement with stability control to be
completed.
Set the nominal buffer value for the measured temperature with
<
▲><▼>.
8
Take over the calibration value for the asymmetry with
<ENTER>.
The calibration display for the next buffer appears.
9
If necessary, terminate the calibration as a single-point
calibration with <M>.
The calibration record is displayed.
pH
7.000
24.8 °C
ConCal
1
22.09.2008
08:00
Asymmetry
pH
2
7.000
24.8 °C
ConCal
22.09.2008
08:00