6 reset, 1 resetting the measurement settings, Reset 4.6.1 – Xylem PH3210 FIELD PH METER User Manual
Page 53: Resetting the measurement settings

pH 3210
Operation
53
ba75793e02
12/2008
4.6
Reset
You can reset (initialize) all sensor settings and sensor-independent
settings separately from each other.
4.6.1
Resetting the measurement settings
Note
The calibration data are reset to the default settings together with the
measuring parameters. Recalibrate after performing a reset.
pH
The following settings for pH measurements are reset to the default
settings with the Reset function:
The sensor settings are reset under the Reset menu item in the
measuring menu. To open the settings, activate the relevant measuring
window in the measured value display and press the <F1>/[Menu] key
shortly.
Setting
Default settings
Buffer AutoCalTEC
Cal. interval
7 d
Unit for slope
mV/pH
Measured parameter
pH
Resolution pH
0.001
Resolution mV
0.1
Asymmetry
0 mV
Slope
-59.16 mV
Man. temperature
25 °C
One point calibration
Off