What to do if...

59

ba75793e02

12/2008

Obviously incorrect

measured values

Meter does not react to

keystroke

You want to know which

software

version is in the meter

Cause

Remedy

pH combination electrode:

– pH combination electrode

unsuitable

– Use suitable combination

electrode

– Temperature difference

between buffer and test sample
too high

– Adjust temperature of buffer

or sample solutions

– Measurement procedure not

suitable

– Follow special procedure

Cause

Remedy

– Operating condition undefined

or EMC load unallowed

– Processor reset:

Press the <ENTER> and
<On/Off> key
simultaneously

Cause

Remedy

– E. g., a question by the service

department

– Switch on the meter.

Open the menu,
<F1

_

>[Menu] / Storage &

config / System / Service
information
. The instrument
data is displayed.