pH 3210

Operation

41

ba75793e02

12/2008

Note
For single-point calibration, the instrument uses the Nernst slope
(-59.16 mV/pH at 25 °C) and determines the zero point of the
electrode.

Continuing with two-

point calibration

(Buffer: TEC)

6

Wait for the end of the measurement with stability control or
accept the calibration value with <ENTER>.
The calibration display for the next buffer appears.

7

If necessary, terminate the calibration as a single-point
calibration with <M>.
The calibration record is displayed.

pH

181.3

mV

24.8 °C

AutoCal

4.010

22.09.2008

08:00

AR

TEC

pH

2

181.3

mV

24.8 °C

AutoCal

22.09.2008

08:00

TEC

8

Thoroughly rinse the combination electrode with distilled water.

9

Immerse the electrode in buffer solution 2.

10

When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <

▲><▼>.

11

Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).