Ph 2 – Xylem PH3210 FIELD PH METER User Manual
Page 41

pH 3210
Operation
41
ba75793e02
12/2008
Note
For single-point calibration, the instrument uses the Nernst slope
(-59.16 mV/pH at 25 °C) and determines the zero point of the
electrode.
Continuing with two-
point calibration
(Buffer: TEC)
6
Wait for the end of the measurement with stability control or
accept the calibration value with <ENTER>.
The calibration display for the next buffer appears.
7
If necessary, terminate the calibration as a single-point
calibration with <M>.
The calibration record is displayed.
pH
181.3
mV
24.8 °C
AutoCal
4.010
22.09.2008
08:00
AR
TEC
pH
2
181.3
mV
24.8 °C
AutoCal
22.09.2008
08:00
TEC
8
Thoroughly rinse the combination electrode with distilled water.
9
Immerse the electrode in buffer solution 2.
10
When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <
▲><▼>.
11
Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).