7 carrying out an automatic calibration (autocal), Carrying out an automatic calibration (autocal) 40, Ph 1 – Xylem PH3210 FIELD PH METER User Manual

Page 40

Advertising
background image

Operation

pH 3210

40

ba75793e02

12/2008

4.4.7

Carrying out an automatic calibration (AutoCal)

Make sure that the correct buffer set is selected in the sensor menu,
Buffer submenu (see page 32).

Of the selected buffer set, use one to five buffer solutions of your choice
in ascending or descending order.

Calibration with Technical buffers is described below (TEC). With other
buffer sets, other nominal buffer values are displayed. Apart from that,
the procedure is identical.

Note
If single-point calibration is set in the menu, the calibration is
automatically terminated after measurement of buffer solution 1 and
the calibration record displayed.

1

In the measured value display, select the measured parameter,
pH or mV with <M>.

2

Start the calibration with <CAL>.
The calibration display appears.

3

Immerse the electrode in buffer solution 1.

4

When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <

▲><▼>.

5

Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).

pH

1

181.3

mV

24.8 °C

AutoCal

22.09.2008

08:00

TEC