Operation

pH 3210

42

ba75793e02

12/2008

Continuing with three-

to five-point calibration

(Buffer TEC)

12

Wait for the measurement with stability control to be completed
or terminate the stability control and take over the calibration
value with <ENTER>.
The calibration display for the next buffer appears.

13

If necessary, terminate the calibration as a two-point calibration
with <M>.
The calibration record is displayed.

pH

1.3

mV

24.8 °C

AutoCal

7.000

22.09.2008

08:00

AR

TEC

pH

3

1.3

mV

24.8 °C

AutoCal

22.09.2008

08:00

TEC

14

Thoroughly rinse the combination electrode with distilled water.

15

Immerse the electrode in buffer solution 3.

16

When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <

▲><▼>.

17

Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).