Xylem PH3210 FIELD PH METER User Manual

Page 46

Advertising
background image

Operation

pH 3210

46

ba75793e02

12/2008

Continuing with two-

point calibration

10

Thoroughly rinse the combination electrode with distilled water.

11

Immerse the electrode in buffer solution 2.

12

When measuring without temperature sensor:
Measure the temperature of the buffer manually and enter it
with <

▲><▼>.

13

Start the measurement with <ENTER>.
The measured value is checked for stability (stability control).

14

Wait for the measurement with stability control to be
completed.
Set the nominal buffer value for the measured temperature with
<

><>.

15

Take over the calibration value for the slope with <ENTER>.
The calibration is completed as a two-point calibration.
The calibration record is displayed.

pH

10.000

24.8 °C

ConCal

2

22.09.2008

08:00

Slope