pH 3210
What to do if...
57
ba75793e02
12/2008
6
What to do if...
Error message
OFL, UFL
Error message
Error
Cause
Remedy
pH combination electrode:
– Measured value outside the
measuring range
– Use suitable combination
electrode
– Air bubble in front of the
junction
– Remove air bubble
– Air in the junction
– Extract air or moisten
junction
– Cable broken
– Replace combination
electrode
– Gel electrolyte dried out
– Replace combination
electrode
Cause
Remedy
pH combination electrode:
– The values determined for zero
point and slope of the
combination electrode are
outside the allowed limits.
– Recalibrate
– Junction contaminated
– Clean junction
– Combination electrode broken
– Replace combination
electrode
Buffer solutions:
– Incorrect buffer solutions
– Change calibration
procedure
– Buffer solutions too old
– Use only once.
Note the shelf life
– Buffer solutions depleted
– Change solutions