VARIO pH
Operation
21
ba75373e04
07/2006
Note
The meter automatically recognizes a temperature sensor. The display
accuracy of the temperature indicates whether or not a temperature sensor
is connected:
Changing the
temperature unit
You can switch the temperature display from °C to °F:
Manual
temperature input
If you have connected an electrode without temperature sensor to the
adapter:
Note
When calibrating without a temperature sensor set the current temperature
of the respective buffer solution manually as well.
Display accuracy of
the temperature (°C or °F)
Temperature sensor
connected?
0,1 yes
1
No
1
Toggle the display between °C and °F with <
§_
> [<Temperature>].
1
Measure the temperature of the measuring solution using a
thermometer.
2
Activate the temperature input with <
§
> [<Temperature>].
The display flashes.
3
Set the temperature of the measuring solution with <
>
> and <
<
>.
4
Confirm the setting with <
g
>.