VARIO pH
What to do if...
45
ba75373e04
07/2006
Display of
l
Instrument does
not react
to keystroke
Sensor symbol
flashes
Message
StoFull
StoFull
StoFull
StoFull
Adapter with electrode
+ measuring solution:
– Conductivity too low
– Use suitable electrode
– Temperature too high
– Use suitable electrode
– Organic liquids
– Use suitable electrode
Cause
Remedy
– Battery almost empty
– Exchange the battery (see
Cause
Remedy
– Operating condition undefined or
EMC load unallowed
– Processor reset:
Press <
1
> and <
8
> at the
same time and release them
again.
Stored data and calibration
values are retained.
Cause
Remedy
– Calibration interval expired
– Recalibrate the measuring
system
Cause
Remedy
– All memory locations are full
– Clear the memory (see section