Displaying the calibration data
The calibration data are displayed after calibrating (see section 4.4.1) and are
accessible at any time.
1
Switch on the meter with <
(
>.
The display test is displayed.
2
During the display test:
Display the date of the last calibration with <
C
>.
3
Press <
g
> to display the asymmetry (ASY).
4
Press <
g
> to display the slope (SLO).
5
Toggle between the display of the slope in mV/pH and % as
necessary with <
"
> [<mV/pH or %>].
The display in % refers to the Nernst slope, 59.2 mV/pH
(100 x the slope determined/Nernst slope).