VARIO pH

Operation

27

ba75373e04

07/2006

4.4.2

Displaying the calibration data

The calibration data are displayed after calibrating (see section 4.4.1) and are
accessible at any time.

1

Switch on the meter with <

(

>.

The display test is displayed.

2

During the display test:
Display the date of the last calibration with <

C

>.

3

Press <

g

> to display the asymmetry (ASY).

4

Press <

g

> to display the slope (SLO).

5

Toggle between the display of the slope in mV/pH and % as
necessary with <

"

> [<mV/pH or %>].

The display in % refers to the Nernst slope, 59.2 mV/pH
(100 x the slope determined/Nernst slope).

6

Switch to the measuring mode with <

g

> or <

E

>.

MEM

CAL

d

C

pH

8

LoBat

min:s

Ident

Lap

mV

h:min

CLR

RESET

LAP

ESC

C A L

5

3 0

Day.Month