9 reset, Reset – Xylem VARIO WATERPROOF PH METER User Manual
Page 38

Operation
VARIO pH
38
ba75373e04
07/2006
4.9
Reset
The following measurement parameters are reset to the delivery status:
Note
The calibration data gets lost when the measurement parameters are reset.
Recalibrate after performing a reset. Stored data are retained.
Resetting the
measurement
parameters
Parameter
Default settings
Measuring mode
pH
Asymmetry
0 mV
Slope -59.2
mV/pH
Calibration procedures
TEC
Temperature, manual
25 °C
1
Using <
C_
>, start the INIT initialization menu.
2
Using <
<
> or <
>
> or <
§
>, select YES for initialization.
3
Confirm with <
g
>.
The reset is performed.
The meter switches to the setting of the calibration interval. The
display flashes. You can change the calibration interval (see section
4.4.3) or switch to the measured value display with <
E
>.