4 storage, Storage – Xylem VARIO WATERPROOF PH METER User Manual

Page 41

Advertising
background image

VARIO pH

Maintenance, cleaning, disposal

41

ba75373e04

07/2006

5.4

Storage

To maintain the operability of the meter and short pH electrode for a long
time, please observe the following instructions:

z

Do not store the short pH electrode for a longer period of time without the
watering cap put on. The short pH electrode might be damaged by this.

z

If possible, leave the short pH electrode or adapter connected to the meter
and avoid to plug it on or off when it is not necessary. Thus, the watertight
plug-in connector will be well protected.

z

To store electrodes for a longer period of time, please follow the
instructions in the operating manual of the electrode.

z

For a longer storage period, moisten the sponge in the watering cap with
a few drops of pH 4 buffer solution and put the watering cap on the short
pH electrode.
During a longer storage period, occasionally check the sponge in the
watering cap for sufficient moisture.

Moistening the

watering cap

Storing the short

pH electrode

In order to store an already used short pH electrode separately from the
meter, protect the electrode with the watering cap (1) and the socket of the
short pH electrode with the original protective cap (2).

pH

4

pH

4

2

1

Advertising