6 what to do if, What to do if, Ee chapter 6 w – Xylem VARIO WATERPROOF PH METER User Manual
Page 43: Chapter 6 w, 6what to do if

VARIO pH
What to do if...
43
ba75373e04
07/2006
6
What to do if...
Error message
0FL
0FL
0FL
0FL
,
UFL
UFL
UFL
UFL
Error message
Err3
Err3
Err3
Err3
Cause
Remedy
– Air bubble in front of the junction
– Remove air bubble
– Electrode defective
– Replace electrode
Cause
Remedy
Erroneous calibration
The data of the electrode are outside
the specified characteristics.
– Recalibrate
– Reset to default setting (see
Electrode:
– Junction contaminated
– Clean junction
– Membrane contaminated
– Clean membrane
– Moisture in the plug
– Dry plug
– Electrode obsolete
– Replace electrode
– Electrode broken
– Replace electrode
Meter:
– Incorrect buffer set
– Select correct buffer set
– Incorrect solution temperature
(without temperature sensor)
– Set the correct temperature
– Socket damp
– Dry socket
Buffer solutions:
– Incorrect buffer set
– Use a suitable buffer set
– Buffer solutions too old
– Use only once.
Note the shelf life