Xylem VARIO WATERPROOF PH METER User Manual
Page 44

What to do if...
VARIO pH
44
ba75373e04
07/2006
Error message
Err4
Err4
Err4
Err4
Error message
Err5
Err5
Err5
Err5
No stable
measured value
Cause
Remedy
No short pH electrode or adapter
– Connect short pH electrode or
adapter
– Check the plug connection
– Switch the meter off and then on
again.
– Perform processor reset
– Replace short pH electrode or
adapter
Cause
Remedy
Short pH electrode or adapter
defective
– Check the plug connection
– Switch the meter off and then on
again.
– Processor reset:
Press <
1
> and <
8
> at the
same time and release them
again.
Stored data and calibration
values are retained.
– Replace short pH electrode or
adapter
Cause
Remedy
Electrode:
– Junction contaminated
– Clean junction
– Membrane contaminated
– Clean membrane
Measuring solution:
– pH value not stable
– Measure with air excluded if
necessary
– Temperature not stable
– Adjust temperature if necessary