Operation

VARIO pH

24

ba75373e04

07/2006

4.4.1

Carrying out calibration

Preparatory

activities

Manual

temperature input

When measuring via an adapter with a sensor without temperature sensor,
the temperature of the buffer solutions has to be input manually during
calibration.

During calibration, the temperature can be input before measuring each
buffer. BUF and a temperature value (without decimal place) is displayed
then.

Calibration
procedure:

1

Switch on the meter with <

(

>.

After the display test, a measured value is displayed.

2

Keep buffer solutions from a TEC or Nist/DIN buffer set ready. For
this purpose: Fill a sufficient quantity of each buffer solution in a small
beaker. The quantity suffices when the minimum immersion depth for
the sensor is granted (see section 4.3.1).

1

Activate the temperature input with <

§

> [<Temperature>].

The selected display flashes.

2

Measure the current temperature of the buffer solution.

3

Enter the temperature of the buffer solution with <

<

> or <

>

>.

4

Confirm the setting with <

g

>.

1

Start the calibration with <

C

>.

The display indicates the adjusted buffer set, TEC BUF or NIST BUF.
The

c

status indicator appears.

2

If necessary, change the adjusted buffer set with <

§

> [<Buffer set>]

(TEC BUF or NIST BUF).

3

Confirm the adjusted buffer set with <

g

>. BUF1 is displayed.

4

If necessary, enter the temperature of the first buffer solution
manually.

5

Immerse the electrode in the first buffer solution.

MEM

pH

8

LoBat Day.Month

min:s

Ident

Lap

mV

h:min

CLR

RESET

LAP

ESC

C

0 0 1

0

2 5

d

CAL

1