1 carrying out calibration, Carrying out calibration – Xylem VARIO WATERPROOF PH METER User Manual
Page 24

Operation
VARIO pH
24
ba75373e04
07/2006
4.4.1
Carrying out calibration
Preparatory
activities
Manual
temperature input
When measuring via an adapter with a sensor without temperature sensor,
the temperature of the buffer solutions has to be input manually during
calibration.
During calibration, the temperature can be input before measuring each
buffer. BUF and a temperature value (without decimal place) is displayed
then.
Calibration
procedure:
1
Switch on the meter with <
(
>.
After the display test, a measured value is displayed.
2
Keep buffer solutions from a TEC or Nist/DIN buffer set ready. For
this purpose: Fill a sufficient quantity of each buffer solution in a small
beaker. The quantity suffices when the minimum immersion depth for
the sensor is granted (see section 4.3.1).
1
Activate the temperature input with <
§
> [<Temperature>].
The selected display flashes.
2
Measure the current temperature of the buffer solution.
3
Enter the temperature of the buffer solution with <
<
> or <
>
>.
4
Confirm the setting with <
g
>.
1
Start the calibration with <
C
>.
The display indicates the adjusted buffer set, TEC BUF or NIST BUF.
The
c
status indicator appears.
2
If necessary, change the adjusted buffer set with <
§
> [<Buffer set>]
(TEC BUF or NIST BUF).
3
Confirm the adjusted buffer set with <
g
>. BUF1 is displayed.
4
If necessary, enter the temperature of the first buffer solution
manually.
5
Immerse the electrode in the first buffer solution.
MEM
pH
8
LoBat Day.Month
min:s
Ident
Lap
mV
h:min
CLR
RESET
LAP
ESC
C
0 0 1
0
2 5
d
CAL
1