2 ph value / orp voltage, 1 general information, Ph value / orp voltage 4.2.1 – Xylem MULTI 3400 MULTI-PARAMETER METERS User Manual

Page 16: General information

Advertising
background image

Operation

Multi 3400i

16

4.2

pH value / ORP voltage

4.2.1

General information

Preparatory activities

Perform the following preparatory activities when you want to measure:

Note
Incorrect calibration of pH electrodes leads to incorrect measured val-
ues. Calibrate regularly before measuring. You can only connect elec-
trodes of the NTC30 type or without temperature sensor.

Warning
When connecting an earthed PC/printer, measurements cannot be per-
formed in earthed media as incorrect values would result. The RS232
interface is not galvanically isolated.

1

Connect the pH electrode to the measuring instrument.
If necessary, press the <M> key repeatedly until the status dis-
play pH (pH measurement) or U (measurement of the ORP
voltage) appears.

2

Adjust the temperature of the buffer solutions or test solutions,
or measure the current temperature, if you measure without a
temperature sensor.

3

Calibrate or check the measuring instrument with the elec-
trode.

4

Using <

> <>, toggle between the pH or mV measuring

modes.