Xylem MULTI 3400 MULTI-PARAMETER METERS User Manual

Page 21

Advertising
background image

Multi 3400i

Operation

21

Sample printout:

Calibration evaluation

After calibrating, the measuring instrument automatically evaluates the
calibration. The asymmetry and slope are evaluated separately. The
worst evaluation appears on the display.

CALIBRATION PROTOCOL
02.03.01 14:19
Device No.: 12345678
Calibration pH
Cal time: 01.03.01 / 15:20
Cal interval: 7d
AutoCal TEC Tauto
Buffer 1 2.00
Buffer 2 4.01
Buffer 3 7.00 *
Buffer 4 10.00
C1 184.1 mV 25.0°C
C2 3.0 mV 25.0°C
S1 -59.4 mV/pH
ASY1 - 4 mV
Probe: +++

Display

Asymmetry
[mV]

Slope
[mV/pH]

-15 ... +15

-60.5 ... -58

-20 ... +20

-58 ... -57

-25 ... +25

-61 ... -60.5
or
-57 ... -56

Clean the electrode according
to the electrode operating
manual

-30 ... +30

-62 ... -61
or
-56 ... -50

E3

Eliminate the error according to
chapter 6 W

HAT

TO

DO

IF

...

< -30 or
> 30

... -62 or
... -50