background image

Multi 3400i

Operation

29

4.3.4

AutoRead AR (Drift control)

The AutoRead function (drift control) checks the stability of the mea-
surement signal. The stability has a considerable impact on the repro-
ducibility of the measured values.

Criteria

With identical measurement conditions, the following applies:

1

Call up the measuring mode with <M> and/or <

> <> .

2

Activate the AutoRead function with <AR>. The current mea-
sured value is frozen (hold function).

3

Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached. This mea-
sured value is transmitted to the interface.

4

If necessary, start the next AutoRead measurement with
<RUN/ENTER>.

5

To terminate the AutoRead function: Press the <AR> key.

8

LoBat

Time

Baud

Day.Month No.

Ident

Year

Sal

%

mbar

1

S

RCL

Store

Auto

Cal

°

F

1/cm

/

K

%

TDS

UpH

Sal

mV/pH

S/

m

cm

cm

M

Lin

Tref25 Tref20

nLF

AutoCal DIN

AutoCal TEC

6

7

1

0 8

O

°

C

mg/l

TP

ARng

AR

7

Measuring mode

Reproducibility

Response time

D. O. concentra-
tion

Better than 0.05 mg/l

> 10 seconds

D. O. saturation in-
dex

Better than 0.6 %

> 10 seconds