7 transmitting data, 1 data transmission interval (int 2), Transmitting data 4.7.1 – Xylem MULTI 3400 MULTI-PARAMETER METERS User Manual

Page 53: Data transmission interval (int 2), T 2 n 1

Advertising
background image

Multi 3400i

Operation

53

4.7

Transmitting data

You have the following possibilities of transmitting data:

z One of the following options:

– With the AutoStore function (page 44), measured values are

periodically saved internally (save interval Int 1) and output on the
interface.

– With the Data transmission interval function (Int 2), measured

values are periodically output on the interface (see below).

AutoStore (Int 1) covers the Data transmission interval

(Int 2).

z Press the <RUN/ENTER> key

This manually triggers a data transmission of the current measured
value to the serial interface at any time - independently of the select-
ed intervals.

z With the Output data storage function (page 46), calibration data or

saved measured values are output on the interface.

4.7.1 Data transmission interval (Int 2)

The interval to the data transmission (Int 2) determines the chronolog-
ical interval between automatic data transmissions. After the selected
interval expires, the current data record is transmitted to the interface.

Setting the

Data transmission

interval

The default setting for the interval is OFF.
To start the data transmission, set an interval (5 s, 10 s, 30 s, 1 min,
5 min, 10 min, 15 min, 30 min, 60 min):

1

Press the <RUN/ENTER> key and hold it down.

2

Press the <RCL> key. Int 2 appears on the display.

3

Set the required interval between the saving procedures with
<

> <>.

Store

Auto

LoBat

Tref25

nLF

/

K

%

RCL

Sal

°

AR

TP

C

M

m

%

mg/l

cm

S/

cm

S

8

cm

1/

°

F

ARng

Cal

Lin

Tref20

Sal

TDS

mbar

O

UpH

mV/pH

AutoCal DIN

AutoCal TEC

Time

Baud

Day.Month No.

Ident

Year

F

F

O

t 2

n

1

Advertising