Calibration evaluation

After the calibration, the measuring instrument automatically evaluates
the current status of the calibration. The evaluation appears on the dis-
play.

Determining the

cell constant

You can determine the cell constant (conductivity measuring cell Tetra-
Con 325) as follows:

Display

Cell constant [cm

-1

]

0.450 ... 0.500 cm

-1

E3

Eliminate the error according to
chapter 6 W

HAT

TO

DO

IF

...

Outside the range
0.450 ... 0.500 cm

-1

1

Press the <CAL> key repeatedly until the status display

CAL CELL appears.

2

Press the <RUN/ENTER> key. The CAL display appears, as
well as
– the current, calibrated cell constant (with sensor symbol on

the display) or

– the fixed cell constant 0.475 1/cm (without sensor symbol

on the display). In this case, the measurement parameters
are initialized (see section 4.9 R

ESET

).

3

Immerse the conductivity measuring cell in the control standard
solution, 0.01 mol/l KCI.

4

Press the <RUN/ENTER>

key.

The AutoRead measurement to determine the cell constant
starts. The AR display indicator flashes until a stable signal is
reached. The cell constant determined is displayed. The mea-
suring instrument automatically stores the cell constants.

AR

AR

RCL

RCL

O

O

UpH

UpH

Sal

Sal

Store

Store

Auto

Auto

AutoCal DIN

AutoCal DIN

AutoCal TEC

AutoCal TEC

LoBat

LoBat

ARng

ARng

Tref20

Tref20

Lin

Lin

nLF

nLF

Sal

Sal

%

%

mg/l

mg/l

mV/pH

mV/pH

cm

cm

°

°

°

°

TP

TP

m

m

S/

S/

cm

cm

F

F

C

C

%

%

/

/

K

K

M

M

TDS

TDS

1

1

S

S

mbar

mbar

Time

Baud

Day.Month No.

Ident

Year

Time

Baud

Day.Month No.

Ident

Year

5

2

0

0

7

7

4

4

A

A

L

L

C

C

cm

cm

1/

1/

Tref25

Tref25

Cal

Cal

Advertising