Xylem MULTI 3400 MULTI-PARAMETER METERS User Manual
Page 72

Technical data
Multi 3400i
72
Oxi measuring ranges
Measuring range/resolu-
tion temperature
Precision of
Oxi (± 1 digit)
Correction functions of
Oxi
Cond measuring ranges
mg/l
%
Measuring range I
Resolution
0 ... 19.99
0.01
0 ... 199.9
0.1
Measuring range II
Resolution
0 ... 90.0l
0.1
0 ... 600
1
°C
Measuring range (Resolution)
0 ... 50.0 (0.1)
Concentration
± 0.5 % of measured value
at ambient temperature of 5 °C ... 30 °C
Saturation
± 0.5 % of measured value
when measuring in the range of
± 10 K around the calibration temperature
Temperature
compensation
< 2 % at 0 ... 40 °C
T [°C]
± 0.1
Salinity correction
0 ... 70.0 SAL
Air pressure correction
automatic through built-in
pressure sensor in the range 500 ... 1100
mbar
Measuring range
Resolution
[µS/cm]
0 ... 1999
1
[mS/cm]
0.00 ... 19.99
0.0 ... 199.9
0 ... 500
0.01
0.1
1
SAL
0.0 ... 70.0 accord-
ing to the IOT table
0.1
T [°C]
− 5.0 ... + 105.0
0.1