4.4.4

AutoRead AR (Drift control)

The AutoRead function (drift control) checks the stability of the mea-
surement signal. The stability has a considerable effect on the repro-
ducibility of the measured value.

Note
The current AutoRead measurement can be terminated at any time
(accepting the current value) by pressing <RUN/ENTER>. You can
only change to another measuring mode after completion of AutoRead.

1

Call up the

or SAL measuring mode with <M> and/or <

>

<

>.

2

Immerse the conductivity measuring cell in the test sample.

3

Activate the AutoRead function with <AR>. The current mea-
sured value is frozen (hold function).

4

Start AutoRead with <RUN/ENTER>.
AR flashes until a stable measured value is reached.
This measured value is transmitted to the interface.

5

If necessary, start the next AutoRead measurement with
<RUN/ENTER>.

6

To terminate AutoRead: Press the <AR> key.