Salt content correction

A salt content correction is required in the oxygen concentration mea-
surement of samples with a salt content of more than 1 g/l. To do this,
you have to enter the salinity equivalent of the test sample (range 0.0 -
70.0) and to switch on the salinity correction.

Note
With the Multi 3400i handheld meter, you can measure the salinity.
How to proceed is described in section 4.4.3 M

EASURING

THE

SALINITY

on page 36.

Entering the salt con-

tent

Note
How to switch on the salt content correction is described on page 27.

1

Determine the salinity of the test sample (any method, see also
section 4.4.3 M

EASURING

THE

SALINITY

on page 36).

2

Press the <CAL> key repeatedly until Sal appears on the dis-
play.

3

Enter the salt content with <

> <>.

4

Switch to the measuring mode with <M>.

°