TM 9-4935-601-14-3&P

SPECIFICATION

The performance limits in this specification are valid with the following conditions:

The instrument must have been calibrated at an ambient temperature between +20

°

C and +30

°

C.

The instrument must have a warm up period of at least 20 minutes.

Table 1-2.

ELECTRICAL CHARACTERISTICS

Characteristics

Performance Requirements

Supplemental Information

Excitation Step Pulse

Cable Dielectric set to AIR

Reflected Rise

0.07 feet (

140 ps)

10% to 90%

2.1 cm for metric

Aberrations

±5% peak during 1st 10 feet (300 cm)
after rise
±0.5% peak beyond 10 feet (300 cm)

Noise Filter "Out"

Jitter

0.02 feet (

40 ps)

Set at X.1 (Cable Dielectric set to

0.6 cm

AIR)

0.1 feet (

200 ps)

Set at X1 (Cable Dielectric

3 cm

set to AIR)

Deflection Factor

5 mp/div to 500 mp/div

7 steps, 1-2-5 sequence

Accuracy

Within ±3%

Gain

At least 3.5:1 from calibrated point

Screwdriver control

Display Noise

±5 mp or less, NOISE FILTER
switch "Out"

Peak

Low Noise Operation

±2 mp or less, NOISE FILTER
switch "in"

Peak

Distance Controls

0 to 2000 feet total

Distance Dial
At X.1 Multiplier
Range

0 to 100 feet
0 to 25 metres for metric

Accuracy

Within ±2% ±0.05 feet

from 2nd to 9th graticule lines

Within ±2% ±0.05 metres for metric

At X1 Multiplier
Range

0 to 1000 feet
0 to 250 metres for metric

Accuracy

Within ±2% ±0.5 feet

from 2nd to 9th graticule lines

Within ±2% ±0.5 metres for metric

REV C FEB 1980