TM 9-4935-601-14-3&P

Adjustment Procedure

Figure 4-10. Pulse Display.

93. Adjust-the front-panel GAIN screwdriver control for a pulse amplitude of 5 divisions. (Adjust the

POSITION controls as necessary.)

94. (a). Set the FEET/DIV (METRES/DIV) to 200 (50).

(b). Check-that the rise of the pulse occurs within 0.1 division horizontally and there is a spike of ap-

proximately 0.2 division at the top corner of the pulse. See Fig. 4-11 for correct and incorrect displays.

95. (a). Adjust-R1543 (LOOP GAIN) located on the MAIN board so that the pulse is as described in the

previous step. See Fig. 4-11 for photos of correct and incorrect displays.

(b). Set the FEET/DIV (METRES/DIV) to 1 (.25), set the MULTIPLIER to.1, and set the CABLE DI

ELECTRIC to OTHER (VAR-fully cw). Use the ZERO REF and POSITION controls as necessary to locate the
pulse at graticule center.

(c). Remove the 50

Ω

terminator from the CABLE connector. Install the GR to bnc adapter and the GR

short. Check that the reflected risetime of the pulse (fall time) is 0.7 division ( 0.875 division). See Fig. 4-12.

Noise

96. Set the mp/DIV to 5, turn the ZERO REF fully counterclockwise and adjust the POSITION controls so

the top of the trace is located on the horizontal centerline.

97. Check-that the peak-to-peak noise is not greater than 10 mp (2 divisions).

98. Push the NOISE FILTER button in and check that the peak-to-peak noise is less than 4 mp (0.8

division).

99. Change the FEET/DIV (METRES/DIV) control to 20 (5). Turn ZERO REF fully clockwise.

100. Use the POSITION controls to set the bottom of the pulse at the horizontal centerline.

101. Check-that the peak-to-peak noise is less than 4 mp.

102. Release the NOISE FILTER button and check that the peak-to-peak noise is less than 10 mp.

103. Adjust--R1657 (LF COMP) located on the MAIN board, for the lowest possible peak-to-peak noise, then

repeat steps 96 through 102 above.

DC Balance

104. Set the mp/DIV to 100.

105. Connect the precision cable from the CABLE connector, to the female adapter, to the 50

Ω

terminator.

Use the POSITION controls to locate the base of the pulse at the horizontal centerline.

REV C FEB 1980

4-15