CALIBRATION PROCEDURE

The first part of this procedure checks the electrical characteristics of the 1502 that appear in the

Specification section of this manual. This procedure can also be used by an incoming inspection facility to
determine acceptability of performance. The second part of this section contains a detailed adjustment
procedure.

It is recommended that these elementary checks be performed before those in the DETAILED CHECKS

AND ADJUSTMENTS section; these checks will identify most conditions calling for troubleshooting and repair.
To aid in troubleshooting, the probable problem area(s) are given after each applicable step. Refer to the Block
Diagram and Schematics in Section 8 as an aid in locating components, test points and circuits.

Table 4-1.

TEST EQUIPMENT REQUIRED

Performance

Description

Requirement

Application

Example

Precision 50

Cable (2)

Known length and

Timing

Tektronix part

impedance

number 012-0482-00

Precision 50

Known impedance

Gain and impedance

Tektronix part

Terminator

checks

number 011-0123-00

BNC female-to-female

Calibration

Tektronix part

adapter

number 103-0028-00

BNC male-to-probe

Calibration

Tektronix part

adapter

number 013-0130-00

BNC-to-GR adapter

Known impedance

Calibration/Performance

Tektronix part

Check

number 017-0064-00

GR short

Low inductance

Calibration/Performance

Tektronix part

Check

number 017-0087-00

Power Module

Calibration

TEKTRONIX TM 503
Power Module

Digital Multimeter

0-200 V dc

Calibration

TEKTRONIX DM 501

Time Mark Generator

5

µ

s to 10

µ

s markers

Calibration

TEKTRONIX TG 501

Pulse Generator

50

µ

s period

Calibration

TEKTRONIX PG 502

2.5

µ

s pulse

Var dc level
5 V pulse

1 X Probe

Calibration

TEKTRONIX P6101

REV B FEB 1980