Figure 2-4 – Atec Tektronix-1502 User Manual

Page 19

Advertising
background image

TM 9-4935-601-14-3&P

16.

RECORD

Two-position lever switch; pushed up and then
released, it initiates X-Y recorder or a chart
recorder.

17.

AC LINE
FUSES

Protection fuses for line power and battery
charging circuits (0.5 A fuses for 115 V ac; 0.3
A fuses for 230 V ac).

PLUG-IN Controls and Connectors

1.

X-Y
OUTPUT

The standard plug-in module for the 1502.
Used to drive an external X-Y Chart

MODULE

Recorder.

X, Y, and
PEN LIFT

Six front panel jacks used for driving an external
X-Y recorder. X jacks are for horizontal drive. Y
jacks are for vertical drive. PEN LIFT jacks are
for pen control.

2.

Y-T CHART
RECORDER

An optional TEKTRONIX Y-T Chart Recorder
which replaces the X-Y OUTPUT MODULE.

OPERATIONAL CHECKOUT

To check the operation of the 1502, follow these step-by-step procedures:

1.

Preset the front panel controls as follows:

FOCUS

Midrange

INTENSITY

Midrange

ZERO REF

Fully cw

POSITION

Midrange

mp/DIV

500

DISTANCE

000

FEET/DIV

1

(METRES/DIV)

(.25)

X1 - X.1

X1

CABLE DIELECTRIC

SOLID POLY

2.

Adjust the INTENSITY and FOCUS controls for a clear

bright trace.

3.

Adjust the POSITION controls to set the trace 2 divisions

below the horizontal centerline.

4.

Attach the precision 50

cable (012-0482-00) to the CABLE connector.

5.

Turn the ZERO REF SET button ccw until the incident pulse edge is located on a vertical reference line.

The incident pulse edge is the initial rise of the step pulse. The vertical reference line may be any line you
choose from the center line to the left side of the crt graticule. We have added an arrow in the second vertical
line to indicate a commonly used reference line.

The reflected pulse from the open end of the 50

cable should appear 3 horizontal divisions to the right

of the reference line in the non-metric version only. The open end of the cable is indicated by the start of a
second rise in the trace (see Fig. 2-4).

6.

Turn the ZERO REF SET control throughout its range to see the incident pulse edge can be set on any

vertical graticule line. Set the incident pulse edge on the vertical reference line.

7.

Set the DISTANCE dial to 050 and check that the top of the step (open cable reflection) is displayed.

8.

Press the ZERO REF CHECK button and check that the incident pulse edge returns to the vertical

reference line of the graticule. Reset the DISTANCE dial to 000.

REV B FEB 1980

Figure 2-4. Incident and Reflected Pulses.

2-4