Figure 4-11 – Atec Tektronix-1502 User Manual

Page 54

Advertising
background image

TM 9-4935-601-14-3&P

Adjustment Procedure

Fig. 4-12. Reflected-Fall Time Check.

106.

Remove the 50 n terminator and check

that the trace shift is less than 1/2 division.

107.

Adjust-R1659(DC BAL) located on the

MAIN board, for the minimum trace shift between
the terminated and unterminated pulse.

Vertical Attenuation

108.

Turn the 1502 top side up on the

bench.

109.

Set the mp/DIV to 500, and turn ZERO

REF fully clockwise.

110.

Set the DM 501 RANGE/FUNCTION

switch to 2 on the DC VOLTS scale.

111.

Connect the lead from the DM 501 LO

terminal to the ground braid of P21 (on the VERT
AMP/SLOW RAMP board).

112.

Connect the lead from the HI terminal

to the left end of R2229. See Fig. 4-13 for the
proper connection points.

113.

Set the 1502 ZERO REF and

POSITION controls so that the DVM reads 1.000

0.008 volt. If the POSITION control cannot be

adjusted to obtain this reading on the DVM, adjust
R1659 (DC BAL) located on the MAIN board, for a
fully counterclockwise setting. Then adjust the
POSITION control for a DVM reading of 1.000

±

0.008 volt.

Figure 4-11. Loop Gain Adjustment.

REV D JUL 1980