Figure 2-3 – Atec Tektronix-1502 User Manual

Page 18

Advertising
background image

TM 9-4935-601-14-3&P

Figure 2-3. 1502 Front Panel Controls.

7.

GAIN

Screwdriver adjust to set the gain

13.

DISTANCE

Indicates the distance from the

of the vertical amplifier.

1502 to the point on the cable
where the display window begins.
Two ranges: 100 feet (25 m) at X.1

8.

NOISE FILTER

Reduces displayed noise. Display

or 1000 feet (250 m) at X1.

rate is reduced by a factor of 10.

Disabled when the FEET/DIV
(METRES/DIV) is at 200 (FIND)
(50).

9.

BATTERY

Meter to indicate the relative
charge of the power pack.

14.

FEET/DIV

Selects the horizontal deflection

10.

ZERO REF

Momentary contact pushbutton.

(METRES/DIV)

factor:

CHECK

When pushed, checks the horizon-

X1 = 1 - 200 ft/div (25 cm -

tal location of the incident pulse on

50 m/div).

the crt when the DISTANCE dial is
being used.

X.1 = 0.1 - 20 ft/div (2.5 cm -
5 m/div).

11.

ZERO REF SET

Horizontal pulse position control
for crt display. Sets the incident
pulse edge to a vertical reference
line of the crt when the DISTANCE

15.

CABLE

Three pushbuttons and a screw-

dial is at 000 or the ZERO REF

DIELECTRIC

driver adjust. Selects the proper

CHECK button is at pushed.

SOLID POLY

velocity of propagation. VAR from
0.55 to 1.0 when the OTHER

SOLID PTFE

pushbutton is pressed. Fully CW is

12.

MULTIPLIER

Two-position switch (red control)

for air dielectric. VAR control has

for X.1 or X1 multiplier. Affects

OTHER VAR

reference marks every 30° to in-

both the DISTANCE dial and the

dicate relative propagation con-

FEET/DIV (M ETRES/DIV) control.

stants.

REV B FEB 1980

2-3